Calcium: Difference between revisions

From Bioblast
No edit summary
No edit summary
Line 26: Line 26:
Please have a look at [[O2k-Publications: Calcium]]. Ā 
Please have a look at [[O2k-Publications: Calcium]]. Ā 


Ā 
Another good starting pointĀ  (and not only for a fluorescence based approach but also for measurements based on ISE) are the web pages of Invitrogen (see main page forĀ  links)and publications cited there. Reading existing calcium related publications should give you a feeling for the issues involved. Some of them (the necessity of Ca2+ buffering,Ā  Ca2+ calibration buffers) are discussed in more detail below. Points to be addressedĀ  will be: Ā 
Another good starting pointĀ  (and not only for a fluorescence based approach but also for measurements based on ISE) are the web pages of Invitrogen (see main page forĀ  links)and publications cited there. Reading existing Ca2+ related publications should give you a feeling for the issues involved. Some of them (the necessity of Ca2+ buffering,Ā  Ca2+ calibration buffers) are discussed in more detail below. Points to be addressedĀ  will be: Ā 
* Can the problem be tackled by using one of the fluorophores from Invitrogene's [[Calcium green]] series? Which of them? Ā 
* Can the problem be tackled by one of the fluorophores form Invitrogene's Ca green series? Which of them? Ā 
* What kind of methods are published?
* What kind of methods are published?


WhatĀ  the [[O2k-Fluorescence LED2-Module]] actually does, is to bring the functionality ofĀ  a cuvette based spectrofluormeter (for selected excitationĀ  emissionĀ  wavelengths) to the O2k chamber. So a publication using a cuvette basedĀ  measurement of a Calcium green fluorophore should also be doable in theĀ  O2k with the fluorescence module. In contrast, the fluorescence moduleĀ  is not intended to supply the functions of e.g. a fluorescenceĀ  microscope (with which e.g. it is possible to quantify the fluorescenceĀ  from a single cell - we are always looking an the entire medium in theĀ  chamber, most of it will be OUTSIDE of the mitos/ cells - so typicallyĀ  we are monitoring concentrations outside of the sample).
WhatĀ  the [[O2k-Fluorescence LED2-Module]] actually does, is to bring the functionality ofĀ  a cuvette based spectrofluormeter (for selected excitationĀ  emissionĀ  wavelengths) to the O2k chamber. So a publication using a cuvette basedĀ  measurement of a Calcium green fluorophore should also be doable in theĀ  O2k with the fluorescence module. In contrast, the fluorescence moduleĀ  is not intended to supply the functions of e.g. a fluorescenceĀ  microscope (with which e.g. it is possible to quantify the fluorescenceĀ  from a single cell A cuvette based approach (and therefore also one using the [[O2k-Fluorescence LED2-Module]]) will always observe the entire medium in theĀ  cuvette / chamber, most of it will be OUTSIDE of the mitos/ cells. Therefore, frequently concentrations outside of the sample are monitored and a fluorophore is chosen that does not permeate cell membranes. However, if the background fluorescence from outside the mitochondria is small enough (or constant enough) also change inĀ  fluorophore intensity generated inside intact cells (or mitochondria ?) might be measured. This will require the the use of cell permeable fluorophores and the cell loading techniques described by the fluorophores producers.


IdeallyĀ  initial experiments will follow published methods but with theĀ  additional dimensions of simultaneous measurement of respiration /Ā  control of oxygen concentration. A publication that can be used as aĀ  starting point should:
IdeallyĀ  initial experiments will follow published methods but with theĀ  additional dimensions of simultaneous measurement of respiration /Ā  control of oxygen concentration. A publication that can be used as aĀ  starting point should:
* address the problems you are interested in
* address the problems you are interested in,
* use one of the Ca green fluorophores to do so
* use one of the differentĀ  [[Calcium green]] fluorophores,
* using a cuvette based approach in a (spectro)fluorometer to do so.
* use a cuvette based approach in a (spectro)fluorometer.


WithĀ  a bit more experience you may be able to follow publications that use aĀ  different Ca fluorphore by substituting it with a suitable Ca Green derivative. Ā 
WithĀ  a bit more experience it may be possible to follow publications that use aĀ  different Ca2+ fluorophore by substituting it with a suitable [[Calcium green]] derivative. Ā 




Line 47: Line 46:
== Ca2+ buffering ==
== Ca2+ buffering ==


MitochondriaĀ  are easily damaged even by the low calcium levels resulting fromĀ  impurities in chemicals and preparations. In fact the nativeĀ  intracellular Ca2+ level may be below the total Ca2+ concentration in aĀ  medium introduced as impurities of standard laboratory chemicals. InĀ  respirometric experiments this is taken care of by "buffering away" allĀ  the Ca2+ with e.g. EGTA. But buffering and measuring small differencesĀ  are of course at cross purpose. One strategy I have seen, is first usingĀ  a very weak Ca2+ buffering, just to keep Ca2+ impurities under controlĀ  and then add external Ca2+ in sufficient amounts to "out-titrate" theĀ  buffer. In the subsequent part of the experiment the mitochondria shouldĀ  be exposed to un-physiologicalyl high Ca2+ concentrations anyway.
MitochondriaĀ  are easily damaged even by the low calcium levels resulting fromĀ  impurities in chemicals and preparations. In fact the nativeĀ  intracellular Ca2+ level may be below the total Ca2+ concentration in aĀ  medium introduced as impurities of standard laboratory chemicals. InĀ  respirometric experiments this is taken care of by "buffering away" allĀ  the Ca2+ with e.g. EGTA. But buffering and measuring small differencesĀ  are of course at cross purpose. One strategy is first usingĀ  a very weak Ca2+ buffering, just to keep Ca2+ impurities under controlĀ  and then add external Ca2+ in sufficient amounts to "out-titrate" theĀ  buffer. In the subsequent part of the experiment the mitochondria are exposed to un-physiologically high Ca2+ concentrations anyway.




Line 53: Line 52:
ToĀ  work at physiological Ca<sup>2+</sup> concentrationsĀ  requires usually Ca<sup>2+</sup> buffering by chelatingĀ  agents. The calculation ofĀ  ''c''(Ca<sup>2+</sup>)<sub>free</sub> in suchĀ  solutions is quite complicated and the results depend i.a on ionicĀ  strength, temperature, and - very strongly - on the pH. Tools for suchĀ  calculations are available on-line at Chris Pattons MAXCHELATOR PageĀ  [http://maxchelator.stanford.edu/Ā  maxchelator.stanford.edu] or theĀ  [http://web.stanford.edu/~cpatton/webmaxcS.htmĀ  web version]. ManyĀ  calculations, published or implemented in programs, depend on a singleĀ  set of data compiled by Martell and Smith <ref>Martell AE, SmithĀ  RM (1989) Critical Stability Constants, 1: Amino Acids. PlenumĀ  Press</ref>.
ToĀ  work at physiological Ca<sup>2+</sup> concentrationsĀ  requires usually Ca<sup>2+</sup> buffering by chelatingĀ  agents. The calculation ofĀ  ''c''(Ca<sup>2+</sup>)<sub>free</sub> in suchĀ  solutions is quite complicated and the results depend i.a on ionicĀ  strength, temperature, and - very strongly - on the pH. Tools for suchĀ  calculations are available on-line at Chris Pattons MAXCHELATOR PageĀ  [http://maxchelator.stanford.edu/Ā  maxchelator.stanford.edu] or theĀ  [http://web.stanford.edu/~cpatton/webmaxcS.htmĀ  web version]. ManyĀ  calculations, published or implemented in programs, depend on a singleĀ  set of data compiled by Martell and Smith <ref>Martell AE, SmithĀ  RM (1989) Critical Stability Constants, 1: Amino Acids. PlenumĀ  Press</ref>.
Note that in spite of the availability of easyĀ  to use software tools the calculation of freeĀ  Ca<sup>2+</sup> concentration is not simple. The accuracy ofĀ  these calculations is severely limited by several factors:
Note that in spite of the availability of easyĀ  to use software tools the calculation of freeĀ  Ca<sup>2+</sup> concentration is not simple. The accuracy ofĀ  these calculations is severely limited by several factors:
* Many underlying thermodynamic constants may not be known precisely
* Many underlying thermodynamic constants may not be known precisely.
* Some thermodynamic constants necessary for adequate temperature correction may not be known at all
* Some thermodynamic constants necessary for adequate temperature correction may not be known at all.
* It may be difficult to set the experimental parameters pH and ionic strength with the necessary precision.
* It may be difficult to set the experimental parameters pH and ionic strength with the necessary precision.
SomeĀ  strategies to simplify the necessary calculationsĀ  are presented in <ref name="Tsien 1989"> Tsien R, Pozzan T (1989)Ā  Measurement of cytosolic free Ca2+ with quin2. Methods Enzymol 172: 230-262. </ref>. Ā 
SomeĀ  strategies to simplify the necessary calculationsĀ  are presented in <ref name="Tsien 1989"> Tsien R, Pozzan T (1989)Ā  Measurement of cytosolic free Ca2+ with quin2. Methods Enzymol 172: 230-262. </ref>. Ā 


== Ca2+ calibration ==
== Ca2+ calibration ==
Before deciding on a calibration method (or specific buffers from thisĀ  method) one should carefully access the needs of the project underĀ  consideration. Is an absolute quantification of Ca2+ necessary at all? Which Ca2+ concentrationsĀ  should beĀ  measured? Maybe one wants to measure extracellular Ca2+ levels? - thenĀ  noĀ  Ca2+ bufferedĀ  Ā  bufferedĀ  calibration solution will be necessaryĀ  at all. Is it possible to do a Ca2+ calibration in the medium used forĀ  the experiment, ie. is it possible to perform the experiment in medium veryĀ  similar to the Ca2+ calibration solutions or can the Ca2+ calibration solutions be modified to be very similar to the used medium?
CalibrationĀ  at low (<Ā  1ĀµM)Ā  Ca<sup>2+</sup> levels isĀ  typically done by using a series of Ca<sup>2+</sup>Ā  calibrationĀ  buffers. Each calibration solution typically contains aĀ  Ā  Ca<sup>2+</sup> chelating agent, a pH buffer,Ā  Ā  CaCl<sub>2</sub>, and a salt (KCl) to adjust ionic strength.Ā  Ā  The pH of all solutionsĀ  have to be adjusted very carefully.Ā  ForĀ  calibrations spanning a largeĀ  range (several orders of magnitude) ofĀ  free Ca<sup>2+</sup>Ā  concentrations it may be necessary toĀ  use calibration buffers with veryĀ  different compositions. UnexpectedĀ  features in calibration curve may in this case mainly due to theĀ  difficulties in calculatingĀ  the free Ca<sup>2+</sup>Ā  concentration, leading to different errors for different mediaĀ  compositions.
A comparable simple method for theĀ  preparation of Ca2+ buffers is presented by Tsien et al. <ref name="Tsien 1989" />. The Calcium calibration buffer kit (C3008MP)Ā  available from Life Technologies (former Invitrogen) is based on this publication. The [https://tools.lifetechnologies.com/content/sfs/manuals/CalciumCalibrationBufferKits_PI.pdf manual ]for this calcium calibration kit can be read as a general introduction in performing Ca2+ calibrations following Tsiens approach, even if the required solutions are prepared in the laboratory and not purchased. Other recipes for Ca2+ calibration buffers can be found in the literature cited in the [[Media:Ca_reading_list.pdf|Ca measurement reading list]].Ā 


A method for theĀ  preparation of Ca buffers is presented in <ref name="Tsien 1989" />. The Calcium calibration buffer kit (C3008MP)Ā  available from Life Technologies (former Invitrogen) is based on this publication. Other recipes for Ca2+ calibration buffers can be found in the literature cited in the [[Media:Ca_reading_list.pdf|Ca measurement reading list]].Ā  Before deciding on a calibration method (or specific buffers from this method) one should carefully access the needs of the project under consideration.Ā  First of all, which Ca concentrationsĀ  should be measured? Maybe one wants to measure extracellular Ca2+ levels? - then noĀ  Ca2+ bufferedĀ  Ā  bufferedĀ  calibration solution will be necessary at all. Is it possible to do a Ca2+ calibration in the medium used for the experiment? Is a absolute quantification of Ca2+ necessary at all?
CalibrationĀ  at low (<Ā  1ĀµM)Ā  Ca<sup>2+</sup> levels is typically done by using a series of Ca<sup>2+</sup> calibrationĀ  buffers. Each calibration solution typically contains aĀ  Ca<sup>2+</sup> chelating agent, a pH buffer,Ā  CaCl<sub>2</sub>, and a salt (KCl) to adjust ionic strength.Ā  The pH of all solutionsĀ  have to be adjusted very carefully.Ā  For calibrations spanning a largeĀ  range (several orders of magnitude) of free Ca<sup>2+</sup>Ā  concentrations it may be necessary to use calibration buffers with veryĀ  different compositions. Unexpected features in calibration curve may in this case mainly due to the difficulties in calculatingĀ  the free Ca<sup>2+</sup> concentration, leading to different errors for different media compositions.





Revision as of 16:56, 23 March 2015


high-resolution terminology - matching measurements at high-resolution


Calcium

Description

Ca2+ is a major signaling molecule in both prokaryotes and eukaryotes. Its cytoplasmic concentration is tightly regulated by transporters in the plasma membrane and in the membranes of various organelles. For this purpose it is either extruded from the cell through exchangers and pumps or stored in organelles such as the endoplasmic reticulum and the mitochondria. Changes in the concentration of the cation regulate numerous enzymes including many involved in ATP utilizing and in ATP generating pathways and thus ultimately control metabolic activity of mitochondria and the of entire cell. Measuring changes in Ca2+ levels is thus of considerable interest in the context of High-resolution respirometry.

Abbreviation: Ca

Reference: O2k-Publications: Calcium


MitoPedia methods: Fluorometry 



under construction !


Ca2+ and HRR

Options for measuring Ca2+

We discuss here measuring Ca2+ concentrations using fluorescence or an ion selective electrode (ISE) system. We recommend to use fluorescence methods for measuring Ca2+ concentrations. Fluorescence based methods for detection of Ca2+ are more easy to set up and are more widely used in biosciences than ISE based methods. However, there may be special applications in which determination of Ca2+ levels via ISE is advantageous. The use of a Ca2+ electrode in mitochondrial research was described by Moreno et al. [1] On this page we discuss topics related to the measurement of Ca2+ that apply to all available methods. Please see Calcium green for specific information about using the fluorophore Ca green to measure Ca2+ concentrations with the O2k-Fluorescence LED2-Module. Please see O2k-TPP+_ISE-Module#Calcium_electrode for specific information about measuring Ca2+ with an ISE system.


How to start a Ca2+ project

Please have a look at O2k-Publications: Calcium.

Another good starting point (and not only for a fluorescence based approach but also for measurements based on ISE) are the web pages of Invitrogen (see main page for links)and publications cited there. Reading existing calcium related publications should give you a feeling for the issues involved. Some of them (the necessity of Ca2+ buffering, Ca2+ calibration buffers) are discussed in more detail below. Points to be addressed will be:

  • Can the problem be tackled by using one of the fluorophores from Invitrogene's Calcium green series? Which of them?
  • What kind of methods are published?

What the O2k-Fluorescence LED2-Module actually does, is to bring the functionality of a cuvette based spectrofluormeter (for selected excitation emission wavelengths) to the O2k chamber. So a publication using a cuvette based measurement of a Calcium green fluorophore should also be doable in the O2k with the fluorescence module. In contrast, the fluorescence module is not intended to supply the functions of e.g. a fluorescence microscope (with which e.g. it is possible to quantify the fluorescence from a single cell A cuvette based approach (and therefore also one using the O2k-Fluorescence LED2-Module) will always observe the entire medium in the cuvette / chamber, most of it will be OUTSIDE of the mitos/ cells. Therefore, frequently concentrations outside of the sample are monitored and a fluorophore is chosen that does not permeate cell membranes. However, if the background fluorescence from outside the mitochondria is small enough (or constant enough) also change in fluorophore intensity generated inside intact cells (or mitochondria ?) might be measured. This will require the the use of cell permeable fluorophores and the cell loading techniques described by the fluorophores producers.

Ideally initial experiments will follow published methods but with the additional dimensions of simultaneous measurement of respiration / control of oxygen concentration. A publication that can be used as a starting point should:

  • address the problems you are interested in,
  • use one of the different Calcium green fluorophores,
  • use a cuvette based approach in a (spectro)fluorometer.

With a bit more experience it may be possible to follow publications that use a different Ca2+ fluorophore by substituting it with a suitable Calcium green derivative.


We have complied a short reading list that may be of interest to those planing to to do Ca2+ measurements, with special emphasis on references describing the preparation of Ca2+ calibration buffers. Ca measurement reading list


Ca2+ buffering

Mitochondria are easily damaged even by the low calcium levels resulting from impurities in chemicals and preparations. In fact the native intracellular Ca2+ level may be below the total Ca2+ concentration in a medium introduced as impurities of standard laboratory chemicals. In respirometric experiments this is taken care of by "buffering away" all the Ca2+ with e.g. EGTA. But buffering and measuring small differences are of course at cross purpose. One strategy is first using a very weak Ca2+ buffering, just to keep Ca2+ impurities under control and then add external Ca2+ in sufficient amounts to "out-titrate" the buffer. In the subsequent part of the experiment the mitochondria are exposed to un-physiologically high Ca2+ concentrations anyway.


Calculation of free Ca2+ concentrations

To work at physiological Ca2+ concentrations requires usually Ca2+ buffering by chelating agents. The calculation of c(Ca2+)free in such solutions is quite complicated and the results depend i.a on ionic strength, temperature, and - very strongly - on the pH. Tools for such calculations are available on-line at Chris Pattons MAXCHELATOR Page maxchelator.stanford.edu or the web version. Many calculations, published or implemented in programs, depend on a single set of data compiled by Martell and Smith [2]. Note that in spite of the availability of easy to use software tools the calculation of free Ca2+ concentration is not simple. The accuracy of these calculations is severely limited by several factors:

  • Many underlying thermodynamic constants may not be known precisely.
  • Some thermodynamic constants necessary for adequate temperature correction may not be known at all.
  • It may be difficult to set the experimental parameters pH and ionic strength with the necessary precision.

Some strategies to simplify the necessary calculations are presented in [3].

Ca2+ calibration

Before deciding on a calibration method (or specific buffers from this method) one should carefully access the needs of the project under consideration. Is an absolute quantification of Ca2+ necessary at all? Which Ca2+ concentrations should be measured? Maybe one wants to measure extracellular Ca2+ levels? - then no Ca2+ buffered buffered calibration solution will be necessary at all. Is it possible to do a Ca2+ calibration in the medium used for the experiment, ie. is it possible to perform the experiment in medium very similar to the Ca2+ calibration solutions or can the Ca2+ calibration solutions be modified to be very similar to the used medium?

Calibration at low (< 1ĀµM) Ca2+ levels is typically done by using a series of Ca2+ calibration buffers. Each calibration solution typically contains a Ca2+ chelating agent, a pH buffer, CaCl2, and a salt (KCl) to adjust ionic strength. The pH of all solutions have to be adjusted very carefully. For calibrations spanning a large range (several orders of magnitude) of free Ca2+ concentrations it may be necessary to use calibration buffers with very different compositions. Unexpected features in calibration curve may in this case mainly due to the difficulties in calculating the free Ca2+ concentration, leading to different errors for different media compositions.

A comparable simple method for the preparation of Ca2+ buffers is presented by Tsien et al. [3]. The Calcium calibration buffer kit (C3008MP) available from Life Technologies (former Invitrogen) is based on this publication. The manual for this calcium calibration kit can be read as a general introduction in performing Ca2+ calibrations following Tsiens approach, even if the required solutions are prepared in the laboratory and not purchased. Other recipes for Ca2+ calibration buffers can be found in the literature cited in the Ca measurement reading list.


Application in biological experiments

Please help us in our instrumental development by explaining what you would like to do/ see in a Ca2+ experiment and what you expect from the measuring method: What Ca concentrations do you want to measure, Ca2+ release or uptake? What total concentration change in a 2 ml chamber (!) do you expect? Do you want to measure in a Ca buffered medium? (decreased sensitivity to changes) or without Ca2+ buffering but then how do you get to physiological Ca2+ concentration? Do you want to measure at physiological Ca2+ concentrations?.......

Please add your comments in the Discussion page. Please contact [email protected] to set up an account.

References

  1. ā†‘ Moreno AJM, Vicente JA (2012) Use of a calcium-sensitive electrode for studies on mitochondrial calcium transport. Methods Mol Biol 810: 207-217.
  2. ā†‘ Martell AE, Smith RM (1989) Critical Stability Constants, 1: Amino Acids. Plenum Press
  3. ā†‘ 3.0 3.1 Tsien R, Pozzan T (1989) Measurement of cytosolic free Ca2+ with quin2. Methods Enzymol 172: 230-262.


Cookies help us deliver our services. By using our services, you agree to our use of cookies.