MiPNet14.06 Instrumental O2 background: Difference between revisions
(Undo revision 187764 by Antunes Diana (talk)) |
(Undo revision 187763 by Antunes Diana (talk)) |
||
Line 6: | Line 6: | ||
|year=2018-08-07 | |year=2018-08-07 | ||
|journal=Mitochondr Physiol Network | |journal=Mitochondr Physiol Network | ||
|abstract=Fasching M, Gnaiger E (2018) O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux. Mitochondr Physiol Network 14.6(07):1-14. ย | |abstract='''Fasching M, Gnaiger E (2018) O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux. Mitochondr Physiol Network 14.6(07):1-14.''' | ||
ย | |||
:ยป Product: [[O2k-FluoRespirometer]], [[Oroboros O2k-Catalogue |O2k-Catalogue]] | |||
:ยป MitoPedia: [[Instrumental_background_oxygen_flux| Instrumental background oxygen flux]] | |||
|keywords=Instrumental O2 background | |||
|mipnetlab=AT_Innsbruck_Oroboros | |mipnetlab=AT_Innsbruck_Oroboros | ||
}} | }} | ||
{{MitoPedia without banner | |||
|abbr=BG | |||
|description=The '''instrumental O<sub>2</sub> background test''' is part 2 of O2k Quality Control, performed after [[MiPNet06.03 POS-calibration-SOP |air calibration of the oxygen sensor]] when starting an experimental series, and performed at intervals of several weeks, depending on the rogorosity of quality control required in an experimental programme. | |||
|info= | |||
}} | |||
__TOC__ | |||
{{Technical support integrated}} | |||
::: '''O2k-Protocols SOP''' | |||
:::: Correction for instrumental background oxygen flux is a standard in high-resolution respirometry, automatically performed by DatLab. Background measurements provide a test of instrument function. In the Oroboros O2k, background corrections are usually within a few % of experimental flux over the entire experimental oxygen range. At minimum activities, however, even the small background effects become highly significant and require compliance to standard operating procedures described in this chapter as part of the [[MitoFit Quality Control System]]. This is part 2 of O2k Quality Control. | |||
:ยป O2k Quality Control 1 ยป[[MiPNet06.03 POS-calibration-SOP]]ยซ | |||
{{Labeling | {{Labeling |
Revision as of 11:30, 13 November 2019
MiPNet14.06 Instrumental O2 background
O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux. |
ยป ยป Versions
Oroboros (2018-08-07) Mitochondr Physiol Network
Abstract: Fasching M, Gnaiger E (2018) O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux. Mitochondr Physiol Network 14.6(07):1-14.
- ยป Product: O2k-FluoRespirometer, O2k-Catalogue
- ยป MitoPedia: Instrumental background oxygen flux
โข Keywords: Instrumental O2 background
โข O2k-Network Lab: AT_Innsbruck_Oroboros
Description
The instrumental O2 background test is part 2 of O2k Quality Control, performed after air calibration of the oxygen sensor when starting an experimental series, and performed at intervals of several weeks, depending on the rogorosity of quality control required in an experimental programme.
Abbreviation: BG
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
- O2k-Protocols SOP
- Correction for instrumental background oxygen flux is a standard in high-resolution respirometry, automatically performed by DatLab. Background measurements provide a test of instrument function. In the Oroboros O2k, background corrections are usually within a few % of experimental flux over the entire experimental oxygen range. At minimum activities, however, even the small background effects become highly significant and require compliance to standard operating procedures described in this chapter as part of the MitoFit Quality Control System. This is part 2 of O2k Quality Control.
- O2k-Protocols SOP
- ยป O2k Quality Control 1 ยปMiPNet06.03 POS-calibration-SOPยซ
Labels: MiParea: Respiration, Instruments;methods
HRR: Oxygraph-2k, O2k-Manual
DatLab, O2k-SOP