| == Leaky Chamber ==
| |
|
| |
|
| === Leaky chamber test ===
| |
| :::* When the leakage is so little that cannot be noticed in a short amount of time, we recommend to follow the ''Leaky chamber test'' below:
| |
|
| |
| ::::# Decrease the oxygen concentration down to ~ 50 ΞΌM (using [[Dithionite]] or N<sub>2</sub> injection)
| |
| ::::# Let the O2k record overnight with the [[closed chamber|chamber closed]] and illumination switched off
| |
| ::::# The oxygen concentration should decrease overtime due to the O<sub>2</sub> consumed by [[OroboPOS]] as it is shown in the figure. In this case, the chamber is assembledΒ properly.
| |
| ::::# If the oxygen concentration increases over time, a reassemble of the chamber is needed.
| |
| [[File:Leaking_chamber_2020-02-03 P2-03.png|1500px]]
| |
|
| |
|
| |
|
| |
| ::::*'''Text in discussion for the page:'''
| |
|
| |
| ::::* To exclude the existence of an unnoticed leak from the chamber, we recommend to follow the ''Leaky chamber test'' below - especially after chamber reassembly:
| |
|
| |
| ::::# Decrease the oxygen concentration down to ~ 50 ΞΌM (using [[Dithionite]] or N<sub>2</sub> injection).
| |
| ::::# Record the O2k measurement overnight with the [[closed chamber|chamber closed]] and illumination switched off.
| |
| ::::# The oxygen concentration should decrease overtime due to the O<sub>2</sub> consumed by [[OroboPOS]] as it is shown in the figure below. In this case, the chamber is assembledΒ properly.
| |
| ::::# If the oxygen concentration increases over time, a reassemble of the chamber is needed.
| |
|
| |
|
| |
| ::::[[User:CecattoC|CecattoC]] ([[User talk:CecattoC|talk]]) 15:52, 19 March 2020 (CET). So, it should be done every reassemble otherwise nobody will noticed that? How frequent you received cases with this or noticed leaky chamber in our lab?
| |