O2k-Open Support Take care Do you perform a stirrer test during oxygen calibration?
In DatLab 7.4, instrumental DL-Protocols guide you automatically towards performing the stirrer test in the process of air calibration. It shows immediately, if your OroboPOS has a satisfactory response time. Further details in MitoPedia: ยปOxygen calibration - DatLab
O2k-Open Support Warning: loss of data Is the noise of the โ€˜O2 slope negโ€™ acceptable?
If you are not certain that the noise of the โ€˜O2 slope negโ€™ is within a suitable range or you want to confirm that your experimental data will not be affected, see further details: MiPNet06.03 POS-calibration-SOP