Template:Fluorescence dyes and settings: Difference between revisions
From Bioblast
(Created page with "{| class="wikitable" |- ! Application !!Sensor!! Filter set !! O2k output !! Light intensity (Amp polarization voltage) !! Gain !! Comment |- | Amplex UltraRed|Amplexยฎ Ul...") ย |
No edit summary |
||
Line 3: | Line 3: | ||
! Application !!Sensor!! Filter set !! O2k output !! Light intensity (Amp polarization voltage)ย !! Gain !! Comment | ! Application !!Sensor!! Filter set !! O2k output !! Light intensity (Amp polarization voltage)ย !! Gain !! Comment | ||
|- ย | |- ย | ||
| [[Amplex UltraRed| | | [[Amplex UltraRed|Amplexโข UltraRed]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type II]] || 100 - 500 || 1000 || at c(AmR) = 10 ยตM | ||
|- ย | |- ย | ||
| [[TMRM]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type III]] || 200 - 500|| 1000 || at c(TMRM) = 2 ยตM | | [[TMRM]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type III]] || 200 - 500|| 1000 || at c(TMRM) = 2 ยตM |
Revision as of 15:22, 18 November 2021
Application | Sensor | Filter set | O2k output | Light intensity (Amp polarization voltage) | Gain | Comment |
---|---|---|---|---|---|---|
Amplexโข UltraRed | Fluorescence-Sensor Green | AmR | Type II | 100 - 500 | 1000 | at c(AmR) = 10 ยตM |
TMRM | Fluorescence-Sensor Green | AmR | Type III | 200 - 500 | 1000 | at c(TMRM) = 2 ยตM |
Safranin | Fluorescence-Sensor Blue | Saf | Type III | 100 - 200 | 1000 | at c(safranin) = 2 ยตM |
Magnesium Greenโข | Fluorescence-Sensor Blue | MgG / CaG | Type I and II | 100 - 500 | 1000 | at c(Mg Green) = 2 ยตM |
Calcium Greenโข | Fluorescence-Sensor Blue | MgG / CaG | Type I | 300 - 500 | 1000 | at c(Ca Green) = 2 ยตM |