Template:Fluorescence dyes and settings: Difference between revisions

From Bioblast
(Created page with "{| class="wikitable" |- ! Application !!Sensor!! Filter set !! O2k output !! Light intensity (Amp polarization voltage) !! Gain !! Comment |- | Amplex UltraRed|Amplexยฎ Ul...")
ย 
No edit summary
Line 3: Line 3:
! Application !!Sensor!! Filter set !! O2k output !! Light intensity (Amp polarization voltage)ย  !! Gain !! Comment
! Application !!Sensor!! Filter set !! O2k output !! Light intensity (Amp polarization voltage)ย  !! Gain !! Comment
|- ย 
|- ย 
| [[Amplex UltraRed|Amplexยฎ UltraRed]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type II]] || 100 - 500 || 1000 || at c(AmR) = 10 ยตM
| [[Amplex UltraRed|Amplexโ„ข UltraRed]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type II]] || 100 - 500 || 1000 || at c(AmR) = 10 ยตM
|- ย 
|- ย 
| [[TMRM]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type III]] || 200 - 500|| 1000 || at c(TMRM) = 2 ยตM
| [[TMRM]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type III]] || 200 - 500|| 1000 || at c(TMRM) = 2 ยตM

Revision as of 15:22, 18 November 2021

Application Sensor Filter set O2k output Light intensity (Amp polarization voltage) Gain Comment
Amplexโ„ข UltraRed Fluorescence-Sensor Green AmR Type II 100 - 500 1000 at c(AmR) = 10 ยตM
TMRM Fluorescence-Sensor Green AmR Type III 200 - 500 1000 at c(TMRM) = 2 ยตM
Safranin Fluorescence-Sensor Blue Saf Type III 100 - 200 1000 at c(safranin) = 2 ยตM
Magnesium Greenโ„ข Fluorescence-Sensor Blue MgG / CaG Type I and II 100 - 500 1000 at c(Mg Green) = 2 ยตM
Calcium Greenโ„ข Fluorescence-Sensor Blue MgG / CaG Type I 300 - 500 1000 at c(Ca Green) = 2 ยตM

In the O2k-LED2-Module (Series D-G) polarization voltage is used, while in the O2k-Fluo Smart Module (Series H-I) fluorescence intensity. To use the same setting in the O2k-Fluo LED2-Module as in the O2k-Fluo Smart-Module use the following Excel sheet:

Download2.png Download the EXCEL calculation-sheet to adjust fluorescence intensity