::: '''Volume calibration of the [[O2k-Chamber]]'''
::: '''Volume calibration of the [[O2k-Chamber]]'''
::::* Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
::::* Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
::::* Follow the instructions in the O2k-Manual: [[MiPNet19.18A O2k-Series G: Start]]
::::* Follow the instructions in the O2k-Manual: [[MiPNet22.11 O2k-FluoRespirometer manual]]
::::* Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
::::* Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.



Revision as of 11:47, 5 April 2018

                



Volume-Calibration Ring

O2k-Catalogue

Description Volume-Calibration Ring (A or B) PVDF for O2k-Stoppers, white; attached to the stoppers.
Product ID 2432A-01 or 2432B-01
Type O2k, Stopper
Link MiPNet22.11 O2k-FluoRespirometer manual
Image
Volume calibration ring A+B PVDF for Stopper.jpg

Technical support

Volume calibration of the O2k-Chamber
  • Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
  • Follow the instructions in the O2k-Manual: MiPNet22.11 O2k-FluoRespirometer manual
  • Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
Volume calibration ring A+B aluminum for Stopper.jpg

Innovation

Replacing the (white) PVDF Volume-Calibration Ring with the (black) aluminum Volume-Calibration Ring allows for more force applied to tighten the ring, thus reducing the risk of a loose fit and the associated danger of altering the calibrated position.