Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Volume-Calibration Ring"

From Bioblast
Line 1: Line 1:
{{Product
{{Product
|description='''Volume-Calibration Ring''' (A or B) for PVDF or PEEK O2k-Stoppers, white; included in stoppers, e.g. [[Stopper\black PEEK\conical Shaft\central Port]].
|description='''Volume-Calibration Ring''' (A or B) for PVDF or PEEK O2k-Stoppers, white; attached to the [[Stopper\black PEEK\conical Shaft\central Port |stoppers]].
|product id=2432A-01 or 2432B-01
|product id=2432A-01 or 2432B-01
|product type=O2k, Stopper
|product type=O2k, Stopper
|info=[[MiPNet12.06]], [[Volume calibration ring (technical service)]]
|info=[[MiPNet12.06]]
|product image=[[Image:Volume calibration ring A+B for PVDF Stopper.jpg|right|180px]]
|product image=[[Image:Volume calibration ring A+B for PVDF Stopper.jpg|right|180px]]
}}
}}
If several [[O2k]] instruments are used, the chambers may be labelled (in [[DatLab]]) A, B, C, D, ..., but volume calibration rings are available only with labels A or B.
Β 
Β 
== Technical service ==
'''Volume calibration of the [[O2k-Chamber]]'''
* Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
* Follow the instructions in the O2k-Manual: [[MiPNet19.18A O2k-Start]]
* Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
Β 
Β 
{{#set:Technical service=Chamber}}
Β 
[[Category:Technical service]]
Β 
__SHOWFACTBOX__

Revision as of 08:40, 9 May 2015

                



Volume-Calibration Ring

O2k-Catalogue

Description Volume-Calibration Ring (A or B) for PVDF or PEEK O2k-Stoppers, white; attached to the stoppers.
Product ID 2432A-01 or 2432B-01
Type O2k, Stopper
Link MiPNet12.06
Image
Volume calibration ring A+B PVDF for Stopper.jpg


Technical service

Volume calibration of the O2k-Chamber

  • Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
  • Follow the instructions in the O2k-Manual: MiPNet19.18A O2k-Start
  • Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.