Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Volume-Calibration Ring"

From Bioblast
Line 9: Line 9:


== Technical service ==
== Technical service ==
{{Technical service}}


'''Volume calibration of the [[O2k-Chamber]]'''
'''Volume calibration of the [[O2k-Chamber]]'''
* Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
* Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).

Revision as of 08:56, 9 May 2015

                



Volume-Calibration Ring

O2k-Catalogue

Description Volume-Calibration Ring (A or B) for PVDF or PEEK O2k-Stoppers, white; attached to the stoppers.
Product ID 2432A-01 or 2432B-01
Type O2k, Stopper
Link MiPNet19.18A O2k-Start
Image
Volume calibration ring A+B PVDF for Stopper.jpg


Technical service

Volume calibration of the O2k-Chamber

  • Loosen the stopper fixation rings (A) and (B) with the L-shaped Allen key (OroboPOS-Service Kit).
  • Follow the instructions in the O2k-Manual: MiPNet19.18A O2k-Start
  • Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.