Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "O2k-OroboPOS-Holder MitoPedia"

From Bioblast
Line 21: Line 21:
[[File:OroboPOS-Holder004.jpg|centre|200px]]
[[File:OroboPOS-Holder004.jpg|centre|200px]]


* Left picture: brand new OroboPOS-Holder, no damage on the thred, no damage on the edge.
* Left picture: brand new OroboPOS-Holder, no damage on the thread, no damage on the edge.
* Center picture: used OroboPOS-Holder, almost no damage on the thred, the edge is slightly damaged. Β 
* Center picture: used OroboPOS-Holder, almost no damage on the thread, the edge is slightly damaged. Β 
* Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.
* Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.
:::: Replacement of a damaged OroboPOS-Holder is recommended. For ordering spare parts: https://www.oroboros.at/index.php/o2k-products/spares/


::::* Replacement of a damaged OroboPOS-Holder is recommended. For ordering spare parts: [https://www.oroboros.at/index.php/o2k-products/spares/ Spare parts & consumables]
Β  ''See [[Talk:O2k-OroboPOS-Holder MitoPedia|Discussion page]]''
Β  ''See [[Talk:O2k-OroboPOS-Holder MitoPedia|Discussion page]]''



Revision as of 13:39, 5 March 2020


high-resolution terminology - matching measurements at high-resolution


O2k-OroboPOS-Holder MitoPedia

Description

The OroboPOS-Holder (POS-Holder), made from blue POM, is screwed into the copper block of the O2k-Main Unit, guiding the OroboPOS-Connector with the OroboPOS sensor head (POS) to the O2k-Chamber, and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-Chamber with the OroboPOS-Seal Tip. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.

Two units of this item are standard components mounted on the O2k-Main Unit.


Reference: OroboPOS-Holder


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber

  • The OroboPOS-Holder may be damaged:
  1. During chamber assembly, when fixing the chamber in a wrong rotational position with the O2k-Chamber-Holder (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (see center picture below). - Follow the details in the video - Β»Chamber service Insert O2k ChamberΒ«
:::# After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the O2k-Main Unit.
  • We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (see pictures below).
OroboPOS-Holder002.jpg
OroboPOS-Holder003.jpg
OroboPOS-Holder004.jpg
  • Left picture: brand new OroboPOS-Holder, no damage on the thread, no damage on the edge.
  • Center picture: used OroboPOS-Holder, almost no damage on the thread, the edge is slightly damaged.
  • Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.