Difference between revisions of "O2k-OroboPOS-Holder MitoPedia"
From Bioblast
Line 14: | Line 14: | ||
:::* The OroboPOS-Holder may be damaged: | :::* The OroboPOS-Holder may be damaged: | ||
:::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber | ::::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber Holder]] (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (''see center picture below''). - ''Follow the details in the video'' - »[https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]« | ||
::::# After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the [[O2k-Main Unit]]. | |||
:::* We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (''see pictures below''). | |||
::::* We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (''see pictures below''). | |||
[[File:OroboPOS-Holder002.jpg|left|200px]] | [[File:OroboPOS-Holder002.jpg|left|200px]] | ||
[[File:OroboPOS-Holder003.jpg|right|200px]] | [[File:OroboPOS-Holder003.jpg|right|200px]] | ||
Line 27: | Line 29: | ||
::::* Replacement of a damaged OroboPOS-Holder is recommended. For ordering spare parts: [https://www.oroboros.at/index.php/o2k-products/spares/ Spare parts & consumables] | ::::* Replacement of a damaged OroboPOS-Holder is recommended. For ordering spare parts: [https://www.oroboros.at/index.php/o2k-products/spares/ Spare parts & consumables] | ||
''See [[Talk:O2k-OroboPOS-Holder MitoPedia|Discussion page]] | ''See'' [[Talk:O2k-OroboPOS-Holder MitoPedia|Discussion page]] | ||
{{MitoPedia O2k and high-resolution respirometry | {{MitoPedia O2k and high-resolution respirometry | ||
|mitopedia O2k and high-resolution respirometry=O2k hardware | |mitopedia O2k and high-resolution respirometry=O2k hardware | ||
}} | }} |
Revision as of 13:41, 5 March 2020
- high-resolution terminology - matching measurements at high-resolution
O2k-OroboPOS-Holder MitoPedia
Description
The OroboPOS-Holder (POS-Holder), made from blue POM, is screwed into the copper block of the O2k-Main Unit, guiding the OroboPOS-Connector with the OroboPOS sensor head (POS) to the O2k-Chamber, and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-Chamber with the OroboPOS-Seal Tip. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.
Two units of this item are standard components mounted on the O2k-Main Unit.
Reference: OroboPOS-Holder
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber
- The OroboPOS-Holder may be damaged:
- During chamber assembly, when fixing the chamber in a wrong rotational position with the O2k-Chamber Holder (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (see center picture below). - Follow the details in the video - »Chamber service Insert O2k Chamber«
- After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the O2k-Main Unit.
- We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (see pictures below).
- Left picture: brand new OroboPOS-Holder, no damage on the thread, no damage on the edge.
- Center picture: used OroboPOS-Holder, almost no damage on the thread, the edge is slightly damaged.
- Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.
- Replacement of a damaged OroboPOS-Holder is recommended. For ordering spare parts: Spare parts & consumables
See Discussion page