Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "High signal at zero oxygen"

From Bioblast
Line 2: Line 2:
__TOC__
__TOC__
{{MitoPedia without banner
{{MitoPedia without banner
|abbr=''R''0
|abbr=
|description=A '''high signal at zero oxygen''' may be observed during [[zero calibration]]. The following instructions show how to further localize this technical problem. For O2k series A-C, please start with the POS-Connector already connected to the main unit and proceed with step 5.
|description=A '''high signal at zero oxygen''' may be observed during [[zero calibration]] (R0). The following instructions show how to further localize this technical problem.  
|info= [[Zero_calibration | Zero calibration]], [[OroboPOS-Connector_MitoPedia#Cleaning_the_electrical_connections | OroboPOS-Connector]]
|info= [[Zero_calibration | Zero calibration]], [[OroboPOS-Connector_MitoPedia#Cleaning_the_electrical_connections | OroboPOS-Connector]]
}}
}}
== High zero current test==
== High zero current test==
=== O2k-series J and XB ===
=== O2k-series D-I and XA ===
::::1. Unscrew the POS from the POS-Connector and unplug the POS-Connector from the O2k-main unit.
::::1. Unscrew the POS from the POS-Connector and unplug the POS-Connector from the O2k-main unit.
::::2. Switch on the O2k and start DatLab.
::::2. Switch on the O2k and start DatLab.
Line 16: Line 19:
::::8. Record the signal for 10 min.
::::8. Record the signal for 10 min.
::::9. Send us the file.
::::9. Send us the file.
 
=== O2k-series A-C ===
:::: Please start with the POS-Connector already connected to the main unit and proceed with step 5.


:::::References:  
:::::References:  

Revision as of 16:20, 22 August 2023


                  


O2k-Open Support

High signal at zero oxygen



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

A high signal at zero oxygen may be observed during zero calibration (R0). The following instructions show how to further localize this technical problem.


Reference: Zero calibration, OroboPOS-Connector

High zero current test

O2k-series J and XB

O2k-series D-I and XA

1. Unscrew the POS from the POS-Connector and unplug the POS-Connector from the O2k-main unit.
2. Switch on the O2k and start DatLab.
3. Record the signal for 10 min.
4. Connect the POS-Connector to the O2k-main unit without the POS. Set an event in DatLab called "POS-Connector".
5. Record the signal for 10 min.
6. Remove the membrane from the POS and wash the cathode and anode with distilled water.
7. Screw the POS to the POS-Connector. Set an event in DatLab called "POS".
8. Record the signal for 10 min.
9. Send us the file.

O2k-series A-C

Please start with the POS-Connector already connected to the main unit and proceed with step 5.
References:
[1] Cleaning the electrical connections
[2] MiPNet19.18B POS-service
[3] MiPNet06.03 POS-calibration-SOP
[4] MiPNet14.06 Instrumental O2 background
Example: First, the OroboPOS-Connector and the POS were disassembled and disconnected from the main unit. Second, the OroboPOS-Connector was connected to the O2k-Main unit without the OroboPOS. Third, the OroboPOS was connected to the OroboPOS-Connector without membranes. After each step, the O2 raw signal [V] was recorded for 10 min. The marks were set to the O2 raw signal [V] to evaluate the zero current.
High zero current good example.png
High zero current test bad exmaple.png
Figure 1. High zero current test. Upper panel: Zero current is in the range of ±0.00099 V. Lower panel: Zero current is higher than 0.00099 V when the OroboPOS-Connector is connected to the O2k-Main unit and when the OroboPOS is connected to the OroboPOS-Connector, indicating a potential electrical problem with both the OroboPOS-Connector and the OroboPOS.