Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Cover-Slip\black"

From Bioblast
m (Protected "Cover-Slip\black" ([edit=sysop] (indefinite) [move=sysop] (indefinite)))
 
(13 intermediate revisions by 7 users not shown)
Line 1: Line 1:
{{Product
{{Technical support}}
|description='''Cover-Slip\black''' should be placed on top of the O2k-Stoppers to avoid light penetration and disturbance of fluorescence signals. The Cover-Slips do not exert any direct effect on oxygen backdiffusion into the [[O2k-Chamber]].
__TOC__
{{MitoPedia without banner
|abbr=
|description=[[Image:Cover-Slip_black.JPG|180px|right]] A '''Cover-Slip''' should be placed on top of the O2k-Stopper to minimize contamination and evaporation of liquid extruding from the capillary of the stopper. The Cover-Slips do not exert any direct effect on oxygen backdiffusion into the [[O2k-chamber]]. Use the the '''Cover-Slip\black''' to avoid light penetration and disturbance of fluorescence signals and generally for optical measurements in the O2k.


[[O2k-Core]]: 2 in [[O2k-Assembly Kit]].
|info= [https://www.oroboros.at/index.php/product/cover-slipblack/ '''Product details and purchase information''']}}
|product id=24411-01
 
|product type=O2k, O2k-Assembly Kit, Stopper
== Technical support ==
|
::::* Cover-Slips cannot be placed on top of the stoppers in O2k-MultiSensor applications with additional sensors inserted through the stopper, or while [[TIP2k]] needles are inserted into the O2k-chamber.
::::* The black cover-slips replaced the previous perspex Cover-Slips (Product ID 24410-01). They are functionally identical, but the black cover slips take into account the optical properties of the [[O2k-Fluorometer]] and prevent light from penetrating through the capillaries of the stoppers when using optical sensors.
== Additional resources ==
::::» [[MiPNet22.11 O2k-FluoRespirometer manual]]
== Keywords ==
{{Template:Keywords: Chamber volume}}
 
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry
}}
}}
* Cover-Slips cannot be placed on top of the stoppers in O2k-MultiSensor applications with additional sensors inserted through the stopper, or while [[TIP2k]] needles are inserted into the O2k-Chamber.
* The black cover-slips replaced the previous perspex [[Cover-Slip]]s (Product ID 24410-01). They are functionally identical, but the black cover slips are of additional importance when optical sensors are protected from light which is prevented from penetrating throught the capillaries of the stoppers.

Latest revision as of 12:48, 10 May 2021


                  


O2k-Open Support

Cover-Slip\black



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

Cover-Slip black.JPG

A Cover-Slip should be placed on top of the O2k-Stopper to minimize contamination and evaporation of liquid extruding from the capillary of the stopper. The Cover-Slips do not exert any direct effect on oxygen backdiffusion into the O2k-chamber. Use the the Cover-Slip\black to avoid light penetration and disturbance of fluorescence signals and generally for optical measurements in the O2k.


Reference: Product details and purchase information


Technical support

  • Cover-Slips cannot be placed on top of the stoppers in O2k-MultiSensor applications with additional sensors inserted through the stopper, or while TIP2k needles are inserted into the O2k-chamber.
  • The black cover-slips replaced the previous perspex Cover-Slips (Product ID 24410-01). They are functionally identical, but the black cover slips take into account the optical properties of the O2k-Fluorometer and prevent light from penetrating through the capillaries of the stoppers when using optical sensors.

Additional resources