O2k-Fluo Smart-Module

From Bioblast
Revision as of 15:49, 8 August 2023 by Tindle Lisa (talk | contribs)


                  


O2k-Open Support

O2k-Fluo Smart-Module



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

[[Description::The O2k-Fluo Smart-Module is an amperometric add-on module to the O2k-Respirometer, adding a new dimension to high-resolution respirometry. Optical sensors are inserted through the front window of the O2k-glass chambers, for measurement of hydrogen peroxide production (Amplex® UltraRed), ATP production (Magnesium Green™), mt-membrane potential (Safranin, TMRM), Ca2+ (Calcium Green™), and numerous other applications open for O2k-user innovation.]]


Reference: [[Info::Product details and purchase information]]

Scope of application

The scope of applications is defined by the excitation wavelengths of the Fluorescence-Sensor Green and Fluorescence-Sensor Blue and the selection of emission filters.
The O2k-Fluo LED2-Module (O2k series D to G) and O2k-Fluo Smart-Module (O2k series H and higher) are designed for measurements performed on samples placed in a stirred and temperature-controlled chamber. Fluorometric protocols based on measurements in cuvettes of various fluorometers can be used with the O2k-Fluo Modules, and extended by simultaneous measurement of respiration and monitoring or regulation of oxygen concentration. The excitation and emission wavelengths of the fluorophore must be suitable for the filters and sensors of the O2k Fluo-Modules (see table below in the section "Fluorescence dyes and settings")
The fluorescence modules are not designed to fulfill the functions of fluorescence microscopes or flow cytometers.


Select Smart Fluo-Sensors

Switching between different excitation wavelengths and filters is achieved by simply exchanging the Fluo-Sensors. Two types of optical sensors are supplied with different LEDs for fluorescence excitation, and the effective spectra of the LEDs are modified by filters.
Select the Fluo-Sensor and filter set from the table below in the section "Fluorescence dyes and settings". Each Fluo-Sensor is delivered with a mounted filter set:
  1. Fluorescence-Sensor Green and Smart Fluo-Sensor Green: 525 nm (dominant wavelength). The installed Filter-Cap is ready for H2O2 measurement with Amplex® UltraRed or for measurements of mt-membrane potential with TMRM.
  2. Fluorescence-Sensor Blue and Smart Fluo-Sensor Blue: 465 nm (dominant wavelength). The installed Filter-Cap is ready for measurements of mt-membrane potential with safranin or Rhodamine 123. A different filter is used for measurement with Magnesium Green™ or Calcium Green™.

PlayVideo.jpg Exchanging a filter set of the Smart Fluo-Sensor

Fluorescence dyes and settings

The light intensity of the LED (LED-intensity) and the signal amplification (Gain) can be adjusted in a wide range. The table suggests initial values, which can be optimised for specific applications.
  • The settings depend on the concentration of the fluorophore, which vary between different applications. Therefore, only recommendations for specific fluorophore concentrations are given. In the Amplex® UltraRed assay the fluorophore is formed during the experiment.
  • The recommendations apply to experiments at 37 °C. The Fluorescence intensity increases strongly at lower temperatures. Then the light intensity is reduced to avoid off-scale signals.
Application Sensor Filter set O2k output Light intensity (Amp polarization voltage) Gain Comment
AmplexTM UltraRed Fluorescence-Sensor Green or Smart Fluo-Sensor Green AmR Type II 100 - 500 1000 at c(AmR) = 10 µM
TMRM Fluorescence-Sensor Green or Smart Fluo-Sensor Green AmR Type III 100 - 500 1000 at c(TMRM) = 2 µM
Safranin Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue Saf Type III 100 - 250 1000 at c(safranin) = 2 µM
Rhodamine 123 Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue Saf Type III 100 - 250 1000 at c(Rh123) = 1 µM
Magnesium Green™ Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue MgG / CaG Type I and II 100 - 500 1000 at c(Mg Green) = 2 µM
Calcium Green™ Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue MgG / CaG Type I 300 - 500 1000 at c(Ca Green) = 2 µM

In the O2k-LED2-Module (Series D-G) polarization voltage is used, while in the O2k-Fluo Smart Module (Series H-I) fluorescence intensity. To use the same setting in the O2k-Fluo LED2-Module as in the O2k-Fluo Smart-Module use the following Excel sheet:

Download2.png Download the EXCEL calculation-sheet to adjust fluorescence intensity
Download2.png

Mounting a filter cap

Stoppers

Use only black PEEK stoppers in conjunction with fluorometric measurements. If necessary, replace the previous white PVDF stoppers. The black stoppers can be used for all HRR applications. See MiPNet22.11 O2k-FluoRespirometer manual for calibration of the 2-mL O2k-chamber volume, which is identical for PEEK and PVDF stoppers.

Connect Smart Fluo-Sensor to O2k

PlayVideo.jpg Smart Fluo-Sensor assembly

Connect the Fluo-Sensor cable to the Fluo plug of the O2k by inserting the male plug of the cable into the female Fluo plug. The red dot on the male plug faces straight upwards. Each Fluo-Sensor can be used on 2-mL O2k-chambers A or B. The blue frame of the chamber window and the Fluo-Sensor are specially designed to align in a specific rotational position with the cable extending horizontally to left (chamber A) and right (chamber B). The Fluo-Sensor is carefully inserted into the window frame and rotated into final position, leaving no gap between window frame and sensor body.
To remove the Fluo-Sensor, carefully pull out the sensor body with slight back and forth rotations. Do not pull on the cable.

Electronic settings and LED intensity

O2k signal and output

  1. O2k signal: The O2k-Fluo Smart-Module is operated through the amperometric (Amp)-Channel of the O2k, with electric current (ampere [Amp]) as the primary signal.
  2. O2k output: type I, II, or III, or combinations.

DatLab 7.4

O2k-Fluorometry and the TIP2k

  • TIP2k: Our tests indicated that the fluorometric signal is not affected by the TIP2k needle inserted into the 2-mL O2k-chamber.

Troubleshooting

References

O2k-Publications in the MiPMap


» O2k-Publications

»O2k-Publications: Topics

Sort in ascending/descending order by a click on one of the small symbols in squares below.
Default sorting: chronological. Empty fields appear first in ascending order. 

{{#ask: Instrument and method::Oxygraph-2k Instrument and method::O2k-Fluorometer |?Was published in year |?Has title |?Mammal and model=Organism |?Tissue and cell=Tissue;cell |format=broadtable |limit=5000 |offset=0 |sort=Was published in year |order=descending }}




» O2k-Publications: O2k-Fluorometry - Abstracts
Sort in ascending/descending order by a click on one of the small symbols in squares below.

{{#ask: Instrument and method::Oxygraph-2k Instrument and method::O2k-Fluorometer |?Was submitted in year |?Has title |?Mammal and model=Organism |?Tissue and cell=Tissue;cell |format=broadtable |limit=5000 |offset=0 |sort=Was submitted in year |order=descending }}


Additional resources

» O2k-FluoRespirometer
» MiPNet22.11 O2k-FluoRespirometer manual
» O2k-Demo experiments
» O2k-Modules
» O2k-Innovation


Questions.jpg



MitoPedia methods: Fluorometry 


MitoPedia O2k and high-resolution respirometry: O2k hardware, O2k-Open Support, O2k-FluoRespirometry